- Model : PCR-800
- Airflow Velocity : 0.2~0.5m/s,Speed adjustable
- Pre-Filter : 1pcs, Polyester fiber,washable
- HEPA Filter : 1 pcs, Efficiency: > 99.995% at particle size of 0.3μm
- Illumination : ≥500Lux
PCR Cabinet PCR-800
PCR cabinet is safety devices that protect laboratory staff from possible contamination and cross-contamination during PCR experiments. Our PCR cabinets are manufactured to the highest industry standards to ensure a safe and reliable laboratory environment. Each component undergoes stringent quality control checks to ensure consistent and reliable performance.
Advantages of PCR Cabinet
Safety Features:
- Audible and visual alarms for filter and UV lamp life status.
- Front sliding sash interlock system linked with fan and UV lamp operation.
- UV-resistant tempered glass used for front sash and side windows.
- Independent capacitors for host and sockets to prevent current overload.
User-friendly Design:
- Standard shelves with IV bar included.
- Memory function to retain settings in case of power failure.
- HEPA filter and UV life indicators for easy replacement scheduling.
- Dual-sided glass windows for convenient experiment observation.
- Easy-to-replace pre-filter extends HEPA filter lifespan.
- Temperature sensor displays real-time operating temperature.
- Waterproof sockets for small instrument use in the operating area.
PCR Cabinet Specifications:
Model | PCR-800 | |
External Size (WxDxH)(mm) | 710x680x1050 (Without Base Stand) | |
710x680x1680 ( With Base Stand) | ||
Internal Size (WxDxH)(mm) | 610x590x625mm | |
Work Surface Height | 750mm | |
Control System | Display | Microprocessor controller with LCD display, display the operating status of each function |
Alarm | Audio and visual alarm on filter and UV changing | |
Interlock | Front sliding sash interlock with fan and UV lamp | |
Life Indicator | HEPA filter and UV life indicators facilitate a timely replacement plan | |
Airflow System | Airflow Velocity | 0.2~0.5m/s, Speed adjustable |
Air filtration System | Pre-Filter | 1pcs, Polyester fiber, washable |
HEPA Filter | 1 pcs, Efficiency: > 99.995% at particle size of 0.3μm | |
Cleanliness Grade | ISO Class 5 (formerly Class 100) | |
Partial air circulation | Perforations at the back wall of the work zone which are designed to eliminate air turbulence and the possibility of dead air corners in the work zone. | |
Cabinet Construction | Main Body | Electro-Galvanized steel with antimicrobial powder coating |
Work Surface | 304 Stainless steel, an angled front, rounded work surface front edge | |
Front Window | Manual sliding sash: 5 mm UV resistant tempered glass | |
Electrical System | UV Lamp | 10W×2 |
With timing function from 0 to 999 minutes | ||
Interlock with front sliding sash. | ||
Emission of 253.7 nanometers for best sterilization effect. | ||
LED Lamp | 8W×2 | |
Consumption | 149W (Not include socket loading power consumption) | |
Power Supply | AC220V±10%, 50/60Hz; 110V±10%, 60Hz | |
Illumination | ≥500Lux | |
Sound Emission | ≤62dB(A) | |
Standard Accessory | 1.LED Lamp×2; 2.UV Lamp×2; 3.Waterproof Socket×1; 4.Shelves with IV Bar | |
Optional Accessory | 1.Base stand; 2.Gas Tap | |
Net Weight | 80kg (Without base stand), 130kg (With base stand) | |
Gross Weight | 115kg (Without base stand), 165kg (With base stand) | |
Shipping Size (W×D×H)(mm) | 920x855x1190 | |
Shipping Volume | 0.94m3 |
PCR Cabinet Package:
Each PCR cabinet undergoes meticulous assembly steps without any delays:
- Complete wrapping with protective film.
- Reinforced edges to prevent collisions during transport.
- Securely housed in fumigation-free plywood crates and pallets, securely bound with iron straps.
Operating principle of PCR Cabinet
WHY SCITEK
- 36 Months Limited Warranty
- CE & FDA European Standards
- 1-2 Days Handling and 7-30 Days Shipping
FAQ
Q: How to decontaminate a pcr niosafety cabinet
A:
- Turn off UV light and unplug.
- Remove items inside.
- Wipe surfaces with disinfectant.
- Let disinfectant sit.
- Turn on UV light for decontamination.
Q: Different between laminar flow cabinet and pcr workstation
A: Laminar flow cabinets provide a sterile airflow environment for various laboratory procedures, while PCR workstations are specifically designed for PCR procedures, with features such as UV decontamination and UV-resistant materials.